Description
SPECIFICATIONS OF A PYROLYTIC MEDICAL WASTE INCINERATOR
An incinerator represents a complete solution to the problem of disposal of all hospital waste types. The liquid fuel is diesel. The incinerator should meet the environment standards.
N° | Expected Technical specifications |
2.01 |
The incinerator shall be fully self contained.
|
2.02 |
The incinerator plant shall incorporate a control panel for its operation. On the control panel we shall have a visual display of the chamber temperatures. This visual display must control and monitor the incinerator and his performance.
|
2.03 |
Capacity of destruction in weight shall be 60 Kg/h. |
2.04 |
It should be able to operate not less than 10 hours/day |
2.05 |
This incinerator must be able to destruct all combustible wastes produced by hospitals, private clinics, laboratories, institutes, etc…
Design Specification : Types A, B, C, D, and E of medical waste
|
2.06 |
“PYROLYTIC” combustion, by controlling the gasification of waste.
|
2.07 |
The incinerator must avoid the release of black smoke and fine dust, during the loadings.
|
2.08 |
It shall be able to reduce the volume of wastes by 98%.
|
2.09 |
It shall be able to hold emission in the second burn with gas residence of not less than 2 seconds.
|
2.10 |
The incineration should be completely free from visible smoke as well as offensive odours.
|
2.11 |
|
2.12 |
The primary combustion chamber shall be constructed from heavy steel plate and the unit is internally lined with a complete layer of hote face refractory and high efficiency insulation.
|
2.13 |
The Internal diameter of the Chimney at least: Ø 400 and its height: 8 m
|
2.14 |
The Volume of the combustion chamber at least: 1.200 L
|
2.15 |
The Dimension of the door for loading in cm at least: 70×70.
|
2.16 |
Burner operation should be Automatic On/Off Fuel : diesel |
2.17 |
Manufacturer to provide necessary information for the best of the installation
|
2.18 |
The control panel shall have all appropriate audible and visual alarms to ensure safe operation.
|
2.19 |
This incinerator with “PYROLYTIC” combustion must have:
. Thickness : ≥100 mm . Nature: 42% of Al203
Thickness: ≥75 mm Nature: Calcium silicate.
Plate of combustion in Carborundum, avoiding the fixing of glass and slags |
2.20 |
A chamber of post combustion of gases
|
2.21 |
A control box ensuring the complete cycle of combustion. |
2.22 |
Fan:
Controls and regulations: Control box watertight to dust, including:
A switch circuit breaker for each engine (ventilators and burners). A timer with adjustable temporization for the regulation of each burner. A regulator with digital watching for the temperature of combustion. A regulator with digital watching for the temperature of post combustion. Electric box
|
2.23 |
The secondary combustion chamber must destroy the contaminant product and must resist to high temperature(1100 °C)
|
2.24 | Process Filtering system: Scruber to be mentionned as optional |
2.25 | Emission Standards Compliance: BS 3316 or équivalent standard |
2.26 |
Capacity to treat Plastic: Not less than 40% by weight |
2.27 | CE Manufacturing Compliance: BS EN 746-2-1997 |